Journal of Microwave Power
and Electromagnetic Energy (JMPEE) |
|
TITLE |
Electrical Characteristics
of Waveguide Applicators for Medical Applications [PDF] |
AUTHORS |
J. Audet, J.C. Bolomey, C. Pichot, D.D. N'Guyen, M. Robillard, M.Chive and Y. Leroy 1980 15 3 177-186 |
YEAR |
|
VOLUME |
|
ISSUE |
|
PAGES |
|
Abstract This work concerns the electrical
properties of waveguide applicators consisting of flanged rectangular
waveguides filled with a dielectric, used in medical applications (microwave thermography and local hyperthermia). The reflection
coefficient and the near field configuration in lossy
materials were obtained in some cases analytically and in some other cases
numerically. The validity of these methods was verified experimentally. The
study shows that the matching of the applicator and the penetration depth in
a lossy material, such as a living tissue, depend
not only on the tissue characteristics but also on the characteristics of the
applicator itself. |