Journal of Microwave Power
and Electromagnetic Energy (JMPEE) |
|
TITLE |
Heating Characteristics
of the Tripas Hyperthermia System for Deep Seated
Malignancy [PDF] |
AUTHORS |
A. Surowiec and H.I. Bicher 1995 30 3 135-140 |
YEAR |
|
VOLUME |
|
ISSUE |
|
PAGES |
|
Abstract A deep heating
hyperthermia device TRIPAS (a triapplicator system)
consisting of three independent, dielectrically loaded horn applicators
operating in phase at 300 MHz was investigated. The heating characteristics
produced by this hyperthermia system were analyzed by means of thermochromic liquid-crystal cards and a modified CDRH (Center for Devices and Regulatory Health) elliptical
phantom. Both homogenous and inhomogeneous phantoms were used, simulating
high and low permittivity tissues (muscle and lung). These equivalent tissues
were made of polyacrylamide gel. The semiquantitative heating pattern analysis showed a
central heating of 1/3 of maximum heating at a depth of Key Words: Deep hyperthermia,
Phantom studies, RF hyperthermia, Horn applicators. |