Journal of Microwave Power and Electromagnetic Energy (JMPEE)

 

TITLE

Measurement of Moisture Content in Granular Ammonium Phosphate by a Microwave Method [PDF]

AUTHORS

A. Kraszewski, S. Kuliński and K. Checiński

1974

9

4

361-372

YEAR

VOLUME

ISSUE

PAGES

 

Abstract

The attenuation of granular ammonium phosphate in a matched, X-band circular waveguide sensor was used to measure its moisture content. In a waveguide length of 250 mm the slope of the calibration curve (attenuation vs. moisture content) of above 12 dB/% , has been obtained. With high meter stability and the density of granular material nearly constant, this slope gives a measurement accuracy of about  ± 0.10 % , in the range of 0.2 to 1.5 % . The measuring sensor, the properties of the material and sources of measurement error are discussed.